Description
Compact bench top model for ELISA work flow management. Accommodates 35 samples, 18 reagent and 12 control locations.
Suitable for infectious diseases like HIV, HCV, Syphilis, Malaria, Dengue, plus Hormones and Autoimmune.
Single probe for dispensing samples and reagents (if required with tips). Automated well washing, with 16 manifold washer, performed at customized volume and cycle.
Automated ELISA reader for single and duel wavelength reading.
Specification
Dimensions
- Length : 1058mm 3.5ft
- Breadth : 520mm 1.8ft
- Height: : 533mm 1.8ft
- Weight : 65kgs
- Shipping weight : 95kgs
Power Supply
- Voltage : AC 110 – 240 V
- Frequency : 50 Hz / 60 Hz
- Power consumption : 300 W
General Specifications
- Consumables : Yes
- Assay per batch : Up to 6 test
- Number of strips : 30 × 8 = 240 micro wells
- Sample tube size : 12 mm × 75 mm (5ml)
- Reagent bottle : 24 × 20 ml
- Controls & Calibrators : 30 × 1.5 ml
- Location : 41 to 58 can be used for loading reagents in 5mL test tube
- Sample dilution : In test tubes
- Tips capacity : 204 tips
Reader Specifications
- Dynamic range : 0.0 - 4.0 O.D
- Wavelength range : 400 - 800 nm
- Filter slots : 405 nm, 450 nm, 492 nm & 630 nm (3 open positions)
- Reading mode : Single & Dual
- Reading time : 20 sec per strip for single wavelength
- Precision : < 1% C.V at 0.5 to 2.0 O.D
- Accuracy : ± 1 or 2.5% (0.000 to 3.00) whichever is greater
Washer Specifications
- Manifold configuration : 8 for aspiration & 8 for dispensing
- Wash cycle : 1 to 9 (repeatable)
- Residual volume : < 8 μl
- Washer buffer capacity : 3 × 1L
- Low buffer alarm : Yes
- Soak time : 20 to 60 sec
- Dispense pressure : Pre-set
- Rinse function : Yes
- Waste bottle : 1 × 2L
Pi-petting Specifications
- Dispensing mode : Fixed needle or Disposable tips
- Sample tips range : 300 μl (5-200 μl)
- Reagent tip range : 300 μl (10-200 μl)
- Precision sample tip : < 3% CV at any operating volume above 10 μl < 2% CV at 100 μl
- Precision reagent tip : < 1% CV at 100 μl
Process Security
- Liquid level sensing : Yes
- Level sensor system : Conductive
- Clot detection : Yes
- Tip detection : Yes
- Alarms : Yes
Software
- Computer include : Desktop PC running TBM O.S (Windows based
- Controlling software : ELAN 30s
- Work protocols (assay) : Unlimited
- Data Processing : Quantitative & Qualitative
- Levey-Jennings : Yes
- West guard rules : Yes
- Process reporting : Event log + error log
- Automatic error recovery : Yes
- Password access control : Yes
Consumables
- Probe clean solution : AC 110 – 240 V
- Carbonated disposable tips : 50 Hz / 60 Hz
- Reagent tips (optional)